ABBOTT VASCULAR INC. MINI TREK¿; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
|
Back to Search Results |
|
| Model Number |
1012270-15 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/13/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was not returned for analysis.Production record and corrective and preventative actions (capa) reviews were performed and revealed no indication of a product quality issue.Additionally, a query of the complaint handling database for the reported lot revealed there is no indication of a lot specific issue.Based on the information received, the investigation determined that the reported balloon rupture appears to be related to operational context.There was no leak noted during preparation, which suggests a product quality issue did not contribute to the reported difficulties.In this case, it is likely that the balloon interacted with the moderately tortuous and moderately calcified anatomy resulting in the reported balloon rupture during inflation.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
|
| |
|
Event or Problem Description
|
|
It was reported that the procedure was to treat a 90% stenosed de novo lesion in the left main coronary artery with moderate calcification and moderate tortuosity.For pre-dilatation, a 2x15mm mini trek balloon dilatation catheter (bdc) was advanced to the target lesion and the balloon inflated; however, the balloon ruptured during the initial inflation at 6 atmospheres, (atm).Therefore, the bdc was removed from the patient, and another trek bdc was used to continue the procedure.There was no adverse patient effect and no clinically significant delay reported in the procedure.No additional information was provided.
|
| |
|
Search Alerts/Recalls
|
|
|