Intuitive surgical, inc.(isi) did receive a da vinci product to perform failure analysis, and the complaint was confirmed.The harmonic insert was analyzed and found to have a broken blade.The blade broke roughly 0.2035¿ from the base.Broken piece was not returned.Size of missing piece measures approximately 0.43¿ x 0.09¿.Components adjacent to the broken blade do not show damage.Additional observation related to customer complaint: the insert was placed to an in-house generator and failed the energy recognition test, instrument generated message " tighten assembly." the probable root cause is attributed to use conditions.Damaged blades result from blade scratches caused by contact with other instruments or hard metal objects (e.G., staples, clips, etc.) while the instrument is activated.These initial damages can worsen due to the ultrasonic vibrations of the harmonic insert, leading to blade failure.Blade damage may be detected by the generator with a solid tone or an error.The probable root cause of the failed energy recognition test was attributed to the broken blade.The missing patient information in sections a and b was either unknown, unavailable, not provided, or not applicable.The expiration date for section d4 is not applicable.Field d6 is blank because the product is not implantable.Information for the blank fields in section e1 is not available.Fields g5 and g7 are not applicable.Field h10 is blank as there are no related report numbers.
|
|
It was reported that during a da vinci-assisted surgical procedure, the harmonic ace insert instrument could not be recognized.The procedure was completed with no patient harm.Intuitive surgical, inc.(isi) followed up with the site to obtain additional information, however, no further details could be provided regarding the reported event.
|