| Model Number |
DDMB1D1 |
| Medical Device Problem Codes |
Pacing Problem (1439); Inappropriate/Inadequate Shock/Stimulation (1574); Device Sensing Problem (2917)
|
| Health Effect - Clinical Code |
Shock from Patient Lead(s) (3162)
|
| Date of Event |
02/23/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Continuation of d10: 5076-52 lead- implanted: (b)(6) 2020.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that the right ventricular (rv) lead exhibited oversensing and undersensing resulting in possible misclassification of episode type and may have led to additional therapies delivered.The rv lead also exhibited t-wave oversensing (twos), the implantable cardioverter defibrillator (icd) delivered inappropriate anti-tachycardia pacing (atp) and shocked for atrial fibrillation (af) with rapid ventricular response which the device detected and treated as ventricular fibrillation (vf).The lead and device remains in used.No further patient complications have been reported as a result of this event.
|
| |
|