| Model Number |
5071-53 |
| Medical Device Problem Codes |
Signal Artifact/Noise (1036); Failure to Capture (1081); Over-Sensing (1438); Pacing Problem (1439)
|
| Health Effect - Clinical Codes |
Stroke/CVA (1770); Loss of consciousness (2418)
|
| Date of Event |
02/09/2026
|
|
Type of Reportable Event
|
Death
|
|
Event or Problem Description
|
|
It was reported that the right ventricular (rv) epicardial leads exhibited loss of capture on the current electrograms (egm), oversensing with sensed events noted post ventricular blanking, and oversensing of noise resulting in inhibition of pacing.It was noted that the patient had an altered mental status and was unresponsive.It was further reported that approximately a month prior, the cardiac resynchronization therapy defibrillator (crt-d) experienced electromagnetic interference from an external, unknown source.The crt-d remains in use.The crt-d and rv epicardial leads remains in use.No further patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Continuation of d10: dtpa2d1 crt-d implanted: (b)(6) 2023 6721m-50 lead implanted: (b)(6) 2019.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was further reported that the patient was paced below the lower programmed rate due to the right ventricular (rv) leads sensing outside of the blanking period.Additionally, the patient experienced a stroke and ultimately passed away.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|