| Model Number |
Mi1210 SYNCHRONY ST |
| Medical Device Problem Code |
Device Appears to Trigger Rejection (1524)
|
| Health Effect - Clinical Code |
Medical device site infection (4845)
|
| Date of Event |
04/29/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device has been explanted and should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
|
| |
|
Event or Problem Description
|
|
Since the implantation surgery, the user had recurring infection.The device was explanted.
|
| |
|
Event or Problem Description
|
|
Since the implantation surgery, the user had recurring infection.The device was explanted.
|
| |
|
Event or Problem Description
|
|
Since the implantation surgery, the user had recurring infection.The device was explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Conclusion: device investigations did not reveal any device defect or problem which is expected to have been present whilst implanted.Mechanical damages found are attributable to the removal surgery.According to the information received from the field the device was explanted due to re-occurring infection and extrusion of the device through the skin.Review of the device_s sterilization records show that the device has been subject to a valid sterilization process, thus no available information points to the implant being the source of the reported issue.However, the presence of the device might have contributed to its subsequent development.In addition in situ measurements show several channels with fluctuating hi status due to undetermined reasons.This is a final report.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: according to the information received from the field the device was explanted due to re-occurring infection and extrusion of the device through the skin.Review of the devices sterilization records show that the device has been subject to a valid sterilization process, thus no available information points to the implant being the source of the reported issue.However, the presence of the device might have contributed to its subsequent development.A revision surgery was performed in (b)(6)2026 however the issue reoccurred and the device was explanted.In addition, in situ measurements show several channels with fluctuating hi status due to undetermined reasons.The explanted device has not been received for investigation yet.
|
| |
|