Catalog Number TSVTB10
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Osteolysis (2377)
Date of Event 04/13/2026
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Zimvie complaint number (b)(4).A1: patient identifier: unknown / not provided.A4: patient weight: unknown / not provided.D10: concomitant medical product: tsvtb10, imp,tsv,3.7,10,mtx,mg, lot: 1296927.Tsvt4b13, imp,tsv,4.1,13,mtx,mg, lot: 1252054.A summary investigation has been completed for bone loss events recognizing that a definitive root cause cannot be identified due to a wide range of external factors (non-design or manufacturing related), including medical conditions (e.G., diabetes, poor bone quality, etc.) / patient habits (e.G., smoking) and surgical technique.Previously completed investigations for bone loss have not identified any signals indicating potential non-conformances affecting the manufacturing and sterilization processes.Furthermore, the probability of manufacturing or design defects that might lead to bone loss and escaping the available detections has been assessed and found remote and almost nonexistent.Should additional information be received which indicates that the device may have caused or contributed to the event, an additional report will be submitted.
 
Event or Problem Description
It was reported that 3 implants at tooth sites #23, 25 and 26 were removed due to a bone loss.Osseointegrated implants for 6 months that have completely resorbed the buccal bone plate.Procedure was completed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMP,TSV,3.7,10,MTX,MG
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
ZIMVIE US CORP LLC
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
ZIMVIE US CORP LLC
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
mark mashburn
4555 riverside drive
palm beach gardens, FL 33410
5619713230
MDR Report Key25109534
Report Number0001038806-2026-02480
Device Sequence Number15757960
Product Code DZE
UDI-Device Identifier00889024019928
UDI-Public(01)00889024019928(17)300303(10)1296927
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
K101977
Exemption Number5645646
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Dentist
Type of Report Initial
Report Date (Section B) 05/07/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Catalogue NumberTSVTB10
Device Lot Number1296927
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/16/2026
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date05/07/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured03/04/2025
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DENTAL IMPLANTS SEE H10 NARRATIVE
Outcome Attributed to Adverse Event Required Intervention;
Patient Age56 YR
Patient SexMale
Patient EthnicityNon Hispanic
-
-