| Model Number |
EVOLUTPRO-23-US |
| Medical Device Problem Codes |
Gradient Increase (1270); Insufficient Device Problem Information (3190); Device Stenosis (4066)
|
| Health Effect - Clinical Codes |
Aortic Valve Stenosis (1717); Dyspnea (1816); Transvalvular Pressure Gradient Increased (4854)
|
| Date of Event |
04/28/2026
|
|
Type of Reportable Event
|
Death
|
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that approximately 6 years and 5 months following the implant of a transcatheter aortic valve, the valve failed due to severe restenosis and a mean gradient of 86 millimeters of mercury (mmhg).The patient experience shortness of breath and was hospitalized as a result.
|
| |
|
Additional Manufacturer Narrative
|
|
Updated data: b2.B5.H6.Additional codes.Corrected data: a4.B3.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
Additional information was received that a redo transcatheter aortic valve replacement (tavr) was scheduled 3 days after the failed valve was identified; however, the patient died.
|
| |
|
Search Alerts/Recalls
|