• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CURONIX LLC FREEDOM PERIPHERAL NERVE STIMULATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CURONIX LLC FREEDOM PERIPHERAL NERVE STIMULATOR Back to Search Results
Model Number STQ4-RCV-B0
Medical Device Problem Code Migration (4003)
Health Effect - Clinical Code Pain (1994)
Date of Event 04/16/2026
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The loss of therapy/no therapy questionnaire was reviewed for potential causes of the reported issue.Based on this review, never achieving therapy, implanting the device off-label, not achieving paresthesia on the table, not performing an x-ray immediately after the implant procedure to confirm device placement, and inadequate fixation have been ruled out.Additionally, improper surgical technique, using incorrect tools, severe force being applied to the implant, excessive twisting or stretching, the patient having contraindicating conditions, the patient having recently gone through an mri, and the patient having non-curonix devices implanted have been ruled out.The stimulator is used to treat pain.The cause of the loss of therapy and new pain is due to migration.However, the cause of migration is unknown.The investigation findings do not lead to a clear conclusion about the cause of the migration.Therefore, conclusion has been selected as unable to determine root cause.Rates reviewed at the most recent complaint trending meeting do not indicate a significant increase in loss of therapy/no therapy.Loss of therapy/no therapy rates remain acceptably low; thus, a capa is not required at this time.Loss of therapy/no therapy rates will continue to be tracked and trended.
 
Event or Problem Description
The patient reported a loss of therapy and new pain in their ankle joint.An x-ray was performed which confirmed migration of the neurostimulator.An explant procedure was performed on (b)(6) 2026.The pain in the ankle joint has somewhat improved after the explant.No further information is available at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FREEDOM PERIPHERAL NERVE STIMULATOR
Common Device Name
PERIPHERAL NERVE STIMULATOR
Manufacturer (Section D)
CURONIX LLC
1310 park central boulevard s.
pompano beach FL 33064
Manufacturer (Section G)
CURONIX LLC
1310 park central boulevard s.
pompano beach FL 33064
Manufacturer Contact
alicia pagliaro
1310 park central boulevard s.
pompano beach, FL 33064
8009655134
MDR Report Key25113745
Report Number3010676138-2026-00075
Device Sequence Number18704506
Product Code GZF
UDI-Device Identifier00850051034413
UDI-Public(01)00850051034413(17)270301
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K233162
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 05/07/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model NumberSTQ4-RCV-B0
Device Lot NumberSWO250328
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/16/2026
Initial Report FDA Received Date05/07/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured03/17/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age51 YR
Patient SexUnknown
-
-