BOSTON SCIENTIFIC CORPORATION MUSTANG?; STENTS, DRAINS AND DILATORS FOR THE BILIARY DUCTS Back to Search Results
Model Number H74939171070440
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/16/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
D2b: pro code (product code): fge, lit.G4: premarket / 510(k) #: k141521, k141597.
 
Event or Problem Description
It was reported that balloon rupture occurred.The 90% stenosed target lesion was located in the moderately tortuous and severely calcified radial shunt.A 7.0 x 40, 40cm mustang balloon catheter was advanced for dilation.During the procedure, the balloon ruptured upon first inflation at 14 atmospheres for 30 seconds.The device was removed without any problem, and the procedure was completed with a different device.There were no patient complications as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MUSTANG?
Common Device Name
STENTS, DRAINS AND DILATORS FOR THE BILIARY DUCTS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
pmt 741 persiaran cassia selat
bandarcassia, pulau pinan 14110
MY   14110
Manufacturer Contact
mark mischke
4100 hamline ave n
arden hills, MN 55112
7634941632
MDR Report Key25113800
Report Number2124215-2026-24802
Device Sequence Number13332135
Product Code FGE
UDI-Device Identifier08714729793588
UDI-Public(01)08714729793588(17)280421(10)36341246
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K141521
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/07/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberH74939171070440
Device Catalogue NumberH74939171070440
Device Lot Number0036341246
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/16/2026
Initial Report FDA Received Date05/07/2026
Date Device Manufactured04/22/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
GUIDEWIRE - RADIFOCUS GUIDEWIRE.; INFLATION DEVICE - AI-25.; INTRODUCER SHEATH - 5F MEDIKIT SHEATH.
Patient SexUnknown
-
-