MAUDE Adverse Event Report: BALT USA OPTIMA COIL SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BALT USA OPTIMA COIL SYSTEM Back to Search Results
Model Number OPTI0407CSFMX
Medical Device Problem Code Migration (4003)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/02/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Balt usa reference: #(b)(4).An evaluation of the actual complaint sample could not be performed as device was unavailable to be return.Based on the provided information, root cause could not be definitively determined.Lack of device return prevented deeper evaluation of the reported issue.Review of the lot history records did not reveal any in-process or lot-specific issue that could account for the observation.No additional complaints against lot number f220100477 has been made for the same issue.Comprehensive analysis of this failure mode has remained subject to monitoring for any unacceptable increase in trend.The submission of this report or related information to the fda, and its release by fda, does not reflect a conclusion by the party submitting this report of the fda that the report or related information is an admission that the manufacturer, their employees, or the device caused or contributed to the reportable event.
 
Event or Problem Description
It was reported that: "physician deployed the optimax 4x7 in an aneurysm to frame.A very strait tail popped out through the stent struts and into the parent vessel.He then deployed an optima helical 3x6 ss and a tail remnant was sticking out as well.He had to place a 2nd stent to tack it up.The physician said it has happened more than a few times with balt coils." incident report form submitted for this complaint indicates that no patient injury was sustained.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPTIMA COIL SYSTEM
Common Device Name
OPTIMA
Manufacturer (Section D)
BALT USA
29 parker
irvine CA 92618
Manufacturer Contact
david vu
29 parker
irvine, CA 92618
9497881443
MDR Report Key25113971
Report Number3014162263-2026-00047
Device Sequence Number21397117
Product Code HCG
UDI-Device Identifier00818053025105
UDI-Public(01)00818053025105(11)220125(17)270125(10)F220100477
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K200030
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/07/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberOPTI0407CSFMX
Device Lot NumberF220100477
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/07/2026
Initial Report FDA Received Date05/07/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured01/25/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Other;
Patient SexUnknown
-
-