• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CONVATEC DOMINICAN REPUBLIC INC DUODERM; DRESSING,WOUND,OCCLUSIVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CONVATEC DOMINICAN REPUBLIC INC DUODERM; DRESSING,WOUND,OCCLUSIVE Back to Search Results
Model Number 187660
Medical Device Problem Codes Delivered as Unsterile Product (1421); Contamination /Decontamination Problem (2895)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device 18 of 25.E1: complainant street address: (b)(6).Complainant country: peru.Based on the available information, this event is deemed to be a reportable malfunction.Photograph, video and/or physical sample evaluation: ¿ there were photographs associated with this case and in these, the reported defect can be seen.No unused return sample was expected.Batch record revision results: lot 5f02076 was manufactured on 10/jun/2025, in bodolay line, with a total of (b)(4) market units (mkus).The complaint engineer performed a batch record review on 29/apr/2026, to verify if all the applicable procedures were followed and no issues were found; all the components for assembly were correct per bill of materials (bom) and all the tooling information documented was also correct, system application product (sap) material identification (id) 1704761 and manufacturing order (b)(4).Review of the batch record showed that all relevant tests required during the manufacturing process and final product release had been fulfilled and met the requirements.No discrepancy related to this issue were found within the documentation.Historical complaints review: on 29/apr/2026, complaints engineer ran a query in database in order to verify the complaints reported for the 5f02076 lot for the malfunction ¿foreign matter within product, sterile products (e.G.Metal contamination in dressings, particulates within fluids, gels etc)¿ defect and no additional complaint was identified.Historical nonconformance review: on 29/apr/2026, complaint engineer ran a query in database looking for any in process nonconformance / corrective action / preventive actions (capa) (s) associated to the malfunction ¿foreign matter within product, sterile products (e.G.Metal contamination in dressings, particulates within fluids, gels etc)¿ for the lot number 5f02076 and as result, no nonconformance / capa (s) for this malfunction were generated during the manufacturing process of the referenced lot.Current quality controls: based on the process instruction (pi), the following tests are performed in the manufacturing lines, in order to identify this failure mode in our manufacturing process: test methods (tm)" package seal integrity nonconformities".Method ¿ 7 (b)(4).Defect rate analysis: there have been 25 defective parts confirmed to date from a lot size of (b)(4) products.This represents a defect rate of only (b)(4) percent, which is well within an appropriate acceptable quality level (aql) (b)(4) percent based on process instruction (pi).In addition, all the in-process testing on this lot did not find a single defective unit, which confirms that the lot is unlikely to breach an aql of (b)(4).Importantly, to date, it is well within our accepted aql level for this type of failure mode or defect.Conclusions: a review of batch records was completed and showed that all relevant tests required during the manufacturing process and final product release had been fulfilled and met the requirements.No discrepancies related to this issue were found within the documentation.This issue will be monitored through the post market product monitoring review process, standard operating procedure (sop).To date no additional information has been received.Should additional information become available, a follow-up report will be submitted.Fda registration number: reporting site: 1049092.Manufacturing site: 9618003.
 
Event or Problem Description
The distributor reported that twenty five of dressings were found with foreign matter inside.The product was not used.Photographs depicting the issue were received from the complainant.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DUODERM
Common Device Name
DRESSING,WOUND,OCCLUSIVE
Manufacturer (Section D)
CONVATEC DOMINICAN REPUBLIC INC
km 18.5 parque industrial
itabo, s.a. haina
san cristobal 91000
DR  91000
Manufacturer (Section G)
CONVATEC DOMINICAN REPUBLIC INC
km 18.5 parque industrial
itabo, s.a. haina
san cristobal 91000
DR   91000
Manufacturer Contact
courtney smith
7815 national service road
suite 600
greensboro, NC 27409
6032470255
MDR Report Key25114982
Report Number9618003-2026-00890
Device Sequence Number10279808
Product Code NAD
UDI-Device Identifier00768455174843
UDI-Public00768455174843
Combination Product (Y/N)N
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 04/17/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number187660
Device Lot Number5F02076
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/17/2026
Initial Report FDA Received Date05/07/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured06/10/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-