| Model Number |
R1040004TP |
| Medical Device Problem Codes |
Circuit Failure (1089); Failure to Calibrate (2440)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/12/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A trilogy 100 ventilator device was returned to a service center for service.There is no allegation of serious or permanent harm or injury.The device was not in use on a patient at the time of the occurrence.During the evaluation, the service technician observed a verify differential pressure sensor calibration test failure and the 200/o2 sensor printed circuit assembly (pca) was replaced to address the failure.Additionally, the handle and front enclosure were replaced due to damage observed, the front panel/keypad led pca was replaced due to the left button responded as pressing twice even though pressing only once, and the base enclosure was replaced due to the screw fastening count exceeded the prescribed value.Perform a repair test, check alarms, and check battery life.Running tests and cleaning were performed to confirm normal operation.Software version 14.2.05 has been confirmed.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported failure of verify differential pressure sensor calibration is accommodated in the product risk management file as being linked to hazard with description patient exposure to function / deterioration of function.Complaints for this failure are reviewed via the post market complaint trending and escalation processes to assess for the observed probability levels and compare them with the predicted levels in the risk file for the hazards linked to the failure.As of the date of submission for this report, there is no indication that complaints for the failure in question has led to unacceptable increase in probability levels for the hazardous situations in scope, and therefore no further action will be pursued.Qa continues to monitor complaints for this issue per the cadence in the complaint trending procedures.The device mentioned in this complaint has exceeded its useful life per the applicable ifu.Multiple service activities occur during the device's useful life according to the device's service manual.These service activities, in addition to any prior complaint investigations, have altered the device from its original state during manufacturing.Any issue found during manufacturing that is potentially documented in the dhr for this device does not have a causal relationship to the allegations or issues in this complaint.Therefore, the dhr for this device will not be reviewed at this time.However, the dhr files for this device are available and can be accessed upon specific request.
|
| |
|
Search Alerts/Recalls
|
|
|