• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESMED PTY LTD ASTRAL 100 - AMER; CONTINUOUS, VENTILATOR, HOME USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESMED PTY LTD ASTRAL 100 - AMER; CONTINUOUS, VENTILATOR, HOME USE Back to Search Results
Model Number 27001
Medical Device Problem Code Protective Measures Problem (3015)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/09/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported to resmed that an astral device displayed an error message (sf179) related to the super capacitor.There was no harm or serious injury reported as a result of this incident.
 
Additional Manufacturer Narrative
The astral device was not returned to resmed.Therefore, resmed is unable to confirm the alleged malfunction.If further information becomes available, a supplemental report will be submitted.Resmed reference: #: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASTRAL 100 - AMER
Common Device Name
CONTINUOUS, VENTILATOR, HOME USE
Manufacturer (Section D)
RESMED PTY LTD
1 elizabeth macarthur drive
bella vista
sydney, nsw 2153
AS  2153
MDR Report Key25115794
Report Number3007573469-2026-01179
Device Sequence Number9292946
Product Code NOU
UDI-Device Identifier00619498270019
UDI-Public(01)00619498270019(11)170821(10)1260755
Combination Product (Y/N)N
Initial Reporter StateKY
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Importer
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 05/07/2026
Report Date (Section F) 05/07/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number27001
Device Catalogue Number27001
Device Lot Number1260755
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date04/09/2026
Device Age103 MO
Event Location Home
Date Report to Manufacturer05/07/2026
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date05/07/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured08/23/2017
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
Date Report Sent to FDA05/07/2026
-
-