• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION TRANSPORT STRETCHER; STRETCHER, WHEELED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION TRANSPORT STRETCHER; STRETCHER, WHEELED Back to Search Results
Model Number P8005H
Medical Device Problem Code Unintended Movement (3026)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/13/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The baxter technician found the four casters worn and needed to be replaced.Per the baxter service manual the bed should be subject to an effective maintenance program.An annual service of the bed is advised in order to maintain its characteristics and performance.Brake casters should be checked for cuts, wear and quality of tread, etc.And replaced when necessary.Apply the brake and make sure the stretcher does not move.If there is movement, look at the brake components for wear.Apply the steer, and make sure the stretcher steers correctly.Look at the steer components for wear.Put the stretcher in neutral.Make sure all four casters rotate and roll freely.Adjust or replace components if necessary.A search of the baxter maintenance records did not show baxter performed any preventative maintenance on this bed.It is unknown if the facility performs preventative maintenance on their beds.The technician replaced the four casters to resolve the reported event.Based on this information, no further action is required.
 
Event or Problem Description
Baxter received a report that std stretcher frame had brakes not holding.This occurred before use.There was no patient/user injury, medical intervention, symptom, or adverse event reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRANSPORT STRETCHER
Common Device Name
STRETCHER, WHEELED
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
HILL-ROM MEXICO
ave.del telefono
no. 200 col. huinala
apodaca, nuevo leon 66640
MX   66640
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242705355
MDR Report Key25118941
Report Number3006697241-2026-00166
Device Sequence Number18704435
Product Code FPO
UDI-Device Identifier00887761000964
UDI-Public(01)00887761000964(11)160531
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
K961437
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 05/07/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberP8005H
Device Catalogue NumberP8005H000821
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/13/2026
Initial Report FDA Received Date05/07/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/31/2016
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-