| Catalog Number |
00504901100 |
| Medical Device Problem Code |
Unsealed Device Packaging (1444)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during stock inspection, the package was found to be nonconforming with a narrow sealing area.There was no patient involvement.Attempts have been made and no further information has been provided.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).G2: foreign - event occurred in japan.-visual evaluation of the returned product found a pouch with peeling in the seal area.The seal was measured and found the seal is non-conforming to zpc 8.400.-review of the device history records identified no deviations or anomalies during manufacturing related to the reported event.-reported event did not occur in an operating room or as part of a medical procedure; medical records are not available for review.-the root cause of the reported event can be attributed to a manufacturing issue.-complaint is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|