The product was returned to pentax medical for repair.Our technician checked the returned unit and confirmed that the cmos module with drive pcb cloudy (not clear view).Based on the result, we concluded that it was caused due to the moisture condensation in the cmos module with drive pcb.In addition, our technician confirmed that the control body fluid damage, the insertion flexible tube buckled, the control body corroded, the mechanical base plate corroded, the operation channel (primary) leak, the operation channel (primary) cut, the remote control buttons cut, the cmos module with drive pcb shadow in image, the led cover glass signs of moisture or fluid under the cover glass; however, these defects are not the main cause, and/or irrelevant to the alleged complaint.Based on the technical report "hr-rpt-0586(image failure)" and/or the risk analysis results, it was evaluated to submit mdr.
|