• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HOYA CORPORATION PENTAX TOKYO OFFICE PENTAX MEDICAL; VIDEO GASTROSCOPE I20C SERIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HOYA CORPORATION PENTAX TOKYO OFFICE PENTAX MEDICAL; VIDEO GASTROSCOPE I20C SERIES Back to Search Results
Model Number EG29-I20C
Medical Device Problem Code No Display/Image (1183)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/16/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The product was returned to pentax medical for repair.Our technician checked the returned unit and confirmed that the cmos module with drive pcb blackout.Based on the result, we concluded that it was caused due to the excessive force applied on the cmos module with drive pcb.In addition, our technician confirmed that the lg connector fluid damage, the pcb for cmos drive fluid damage, and the lg connector corroded; however, these defects are not the main cause, and/or irrelevant to the alleged complaint.Based on the technical report "hr-rpt-0586(image failure)" and/or the risk analysis results, it was evaluated to submit mdr.
 
Event or Problem Description
The time of event is unknown.There was no report of patient harm.Video image failure(blackout ).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENTAX MEDICAL
Common Device Name
VIDEO GASTROSCOPE I20C SERIES
Manufacturer (Section D)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishima-shi, tokyo 196-0 012
JA  196-0012
Manufacturer (Section G)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishima-shi, tokyo 196-0 012
JA   196-0012
Manufacturer Contact
paresh patel
3 paragon drive
montvale, NJ 07645
8004315880
MDR Report Key25120237
Report Number9610877-2026-50969
Device Sequence Number16134659
Product Code FDS
UDI-Device Identifier04961333251858
UDI-Public(01)04961333251858(11)240605
Combination Product (Y/N)N
Initial Reporter StateDC
Initial Reporter CountryUS
PMA/510(K) Number
K231249
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/08/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberEG29-I20C
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/16/2026
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/16/2026
Initial Report FDA Received Date05/08/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/05/2024
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-