| Model Number |
DVPB3D4 |
| Medical Device Problem Codes |
Electromagnetic Interference (1194); Over-Sensing (1438)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/14/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that the implantable cardioverter defibrillator (icd) triggered a false positive right ventricular lead integrity alert (lia) due to sensing integrity counter (sic) and high rate non-sustained episodes due to electromagnetic interference exposure.The alert occurred at the same time the patient applied a transcutaneous electrical nerve stimulator (tens) unit.The rv lead impedance trend was confirmed stable and steady state.The icd remains in use.No patient complications have been reported as a result of this event.
|
| |
|
Search Alerts/Recalls
|