A non-health care professional reported that during intraocular lens (iol) implantation surgery, haptic of iol was broken, and no harm was experienced by the patient.The iol was exchanged with an unknown replacement lens during initial implantation procedure and procedure was completed on the same day.Additional information has been received, and it was updated accordingly.
|
|
Additional information was provided in d.9., h.3., h.6.And h.11.The lens was returned positioned incorrectly in the lens case base (lens case lid was not returned).Solution was dried on the lens.One haptic was broken from the distal area (returned).A device history record review and a non-conformance review of the reported lot number was conducted.The deviation review did not reveal any potential contributing factors to the reported complaint and all corresponding production release specifications defined in the device master record were met.The file indicated the use of a qualified cartridge and handpiece.The associated viscoelastic was not provided.It is unknown if a qualified product was used.The reported broken haptic was observed.All lenses are 100 percent inspected for cosmetic attributes and the damage exhibited by the returned complaint sample would not have met company¿s release criteria.Based on our observation, and the review of manufacturing records, it can be reasonably concluded that the damage is not manufacturing related.Due to the presence of surgical solution and the condition of the returned sample, the damage is most likely related to customer handling.It is unknown if a qualified viscoelastic was used.The ifu (instruction for use) instructs the company foldable iols (intraocular lens) are qualified for use with a company qualified delivery system (handpiece and cartridge) and ophthalmic visco surgical device (ovd) combination.The use of an unqualified combination may cause damage to the lens and potential complications during the implantation process.Using holding forceps, grasp the lens by the optic edge and gently place the lens anterior side up into the back of the ovd-filled cartridge.The lens should be inserted until the optic is a little more than half-way inside the cartridge.Use the holding forceps to gently push down on the lens, verifying that the lens is on the bottom surface of the cartridge.Using holding forceps, take the trailing haptic, and gently fold the haptic onto the anterior side of the optic.Slowly grip or push the optic edge to position the lens as far into the cartridge as the forceps will permit, while ensuring the lens remains on the bottom surface of the cartridge and the trailing haptic remains on the optic.Failure to follow these steps may cause the lens to advance incorrectly, causing delivery issues and/or damage.Follow the section regarding directions for use for information on the maximum allowed time for the iol to stay in the folded condition.Failure to adhere to manufacturer¿s recommendations may result in iol damage.During lens loading and insertion, do not allow the company iol to remain in a folded condition within the selected iol delivery system for more than 3 minutes prior to completing insertion into the capsular bag.The manufacturer internal reference number is: (b)(4).
|