| Model Number |
TJF-Q180V |
| Medical Device Problem Code |
Degraded (1153)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/12/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was observed that during the device evaluation, the duodenovideoscope exhibited worn adhesive around objective lens.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for inspection and the reported failure was confirmed.In addition, the following reportable malfunctions were identified during the device evaluation: worn adhesive around objective lens.A root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction: known inherent risk of device.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|
Search Alerts/Recalls
|