| Model Number |
CD1357-40Q |
| Medical Device Problem Code |
Premature Discharge of Battery (1057)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the patient's implantable cardioverter defibrillator (icd) exhibited battery performance alert.The icd was explanted.The patient's condition is unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
Reported event of battery performance alert was confirmed.Premature battery depletion was confirmed by analysis.A device evaluation was performed, and no sources of high currents were noted.The cause of the premature battery depletion was consistent with li cluster formation.From these analyses, in the absence of a high current draw, it is probable that a lithium cluster-induced short circuit caused premature battery depletion.Li clusters are a known depletion mechanism for these advisory products that have been investigated and associated with a field action in october 2016.
|
| |
|
Search Alerts/Recalls
|