BRIDGES CONSUMER HEALTHCARE THERMACARE MENSTRUAL; HOT OR COLD DISPOSABLE PACK. Back to Search Results
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Codes Pain (1994); Partial thickness (Second Degree) Burn (2694)
Date of Event 04/09/2026
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The root cause cannot be identified.There is limited device specific information provided, no batch number was available for evaluation.Without a batch reference number, a manufacturing and technical evaluation cannot be completed for the wrap involved in this case.No product quality related trend was identified for the subclass adverse event safety request for investigation.Care should be taken when using the device, following all safety and use information as provided with the wrap to avoid the risk of blisters and other skin irritations.The manufacturing operations employ quality control procedures which include in process testing, thermal testing and visual inspection, to ensure packaged product quality.There are pre-identified risk factors that could cause a burn listed in the hazard analysis (b)(4).In addition to these hazards, there are multiple risks that are outside the control of the site.These include things like age, skin condition, medical conditions, device use error and off-label use.The warning labels on our product are used to address these risks and relay the appropriate instructions for use to our customers to avoid burns, blisters and skin irritations.
 
Event or Problem Description
On 19-apr-2026, this spontaneous report was received regarding a female consumer (age not provided) in puerto rico in association with a thermacare menstrual heat wrap.On (b)(6) 2026, the consumer applied a menstrual thermacare heat wrap for an unspecified indication.Medical history, concomitant products, and allergies were not provided.On (b)(6) 2026, after using the product the consumer experienced an immediate reaction.I was clarified that she experienced burns with two blisters that were painful.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THERMACARE MENSTRUAL
Common Device Name
HOT OR COLD DISPOSABLE PACK.
Manufacturer (Section D)
BRIDGES CONSUMER HEALTHCARE
1208 king st.
suite 605
chattanooga TN 37403
Manufacturer (Section G)
ANGELINI
1231 wyandotte dr
albany GA 31705
Manufacturer Contact
tonya blackwell
1231 wyandotte dr
albany, GA 31705
2294463083
MDR Report Key25127288
Report Number3007593958-2026-00022
Device Sequence Number8660883
Product Code IMD
UDI-Device Identifier00305733010396
UDI-Public305733010396
Combination Product (Y/N)N
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 05/04/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/19/2026
Initial Report FDA Received Date05/08/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-