It was reported that a patient underwent a vaginal delivery procedure on (b)(6) 2026 and suture was used.During the procedure, at 8:20 am, after the mother gave birth naturally, the midwife prepared to suture 1-degree laceration perineum.After unpacking the suture packaging, it was found that the needle pull off issue had happened.It was immediately stopped from use and kept.After replacing with another suture, it was used normally.No adverse patient consequences were reported.Additional information was requested.
|
|
Product complaint # (b)(4).H3 photo investigation summary: this is an analysis for a photo submitted to ethicon for evaluation.During the visual analysis, the following was observed: the photos show a winding former labeled vcpb945, the suture in position, and an apparent detachment of the needle.A clear analysis of the end point of failure or the needle hole could not be performed due to the position of the needle.As the device was not returned, an analysis investigation could not be performed.Based on the photo review, the event reported is confirmed, however no conclusion or root cause could be determined.Hands on device analysis may provide the additional evidence necessary to confirm the root cause of the reported event.Because the device was not returned our evaluation is limited.We value the opportunity to fully analyze the device upon its return.As part of ethicon quality process all devices are manufactured, inspected, and released to approved specifications.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.The following information was requested, but unavailable: -did the reported issue contribute to any patient adverse consequences? - when was the needle detached/pulled off from the suture (in the package, during removal from package, during handling prior to use on patient or during use on the patient)? please specify - please provide the source or name of person providing answers to follow-up questions: --received information as following from sales rep via phone today.Please refer to the event description, other information requested is unknown.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
|