| Model Number |
CYF-5 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/02/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for inspection.A root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction: component failure, degradation problem identified.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|
Event or Problem Description
|
|
The cystonephrofiberscope was returned to the service center with a report of multiple dark spots on the lens.This does not indicate an mdr reportable event.However, mdr reportable failure was found during testing at the service center.During inspection of the device, detached bending section cover glue was found.There was no patient involvement.
|
| |
|