| Model Number |
TFL-FBX200S |
| Medical Device Problem Code |
Thermal Decomposition of Device (1071)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/05/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for evaluation/inspection.Based on the results of the investigation, the charred marks were identified.It is likely the result of stress; however, a definitive root cause could not be established.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|
Event or Problem Description
|
|
It was observed that during the device evaluation, the single use fiber exhibited charred marks observed on the fractured end.There was no patient involvement.
|
| |
|
Search Alerts/Recalls
|