| Model Number |
S606 |
| Medical Device Problem Codes |
High impedance (1291); Off-Label Use (1494)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/08/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that this implantable cardioverter defibrillator (icd) exhibited high out of range pacing impedances in the right atrial (ra) channel.The patient was noted to have two ra leads, with one connected to the right ventricular (rv) port of the device; however, the health care professional (hcp) did not clarify which ra lead was connected to the rv port.Technical services (ts) was consulted and recommended replacement in the near future.This icd remains in service.No adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
This product investigation is still in progress.This report will be updated when product investigation is complete.
|
| |
|
Additional Manufacturer Narrative
|
|
This product investigation is still in progress.This report will be updated when product investigation is complete.
|
| |
|
Event or Problem Description
|
|
It was reported that this implantable cardioverter defibrillator (icd) exhibited high out of range pacing impedances in the right atrial (ra) channel.The patient was noted to have two ra leads, with one connected to the right ventricular (rv) port of the device; however, the health care professional (hcp) did not clarify which ra lead was connected to the rv port.Technical services (ts) was consulted and recommended replacement in the near future.This icd remains in service.No adverse patient effects were reported additional information was provided stating that this icd was explanted and replaced due to normal battery depletion.Other than the intervention, no additional adverse patient effects were reported.
|
| |
|