| Model Number |
72400025 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
Urinary Incontinence (4572)
|
| Date of Event |
04/13/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device is not available for analysis; therefore, no physical analysis of the product could be performed.Fluid leak and urinary incontinence are acknowledged within the instructions for use (ifu) and are not related to off-label use or failure to follow instructions.The reported patient symptom of urinary incontinence is a known risk associated with this device type and indicated as such in the instructions for use.Based on the information available, a conclusion code of cause not established was assigned to this investigation.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient with an artificial urinary sphincter (aus) experienced recurrent incontinence.During revision surgery, the physician confirmed that the pressure regulating balloon (prb) was empty.The device was explanted and replaced.There were no further patient complications.
|
| |
|