|
A field service engineer (fse) conducted a customer site visit and confirmed the reported issue by reviewing the customer quality control (qc) data.The fse inspected the analyzer and could feel binding in the main arm sampler z axis with the main arm z motor removed.The fse lubricated and adjusted the main arm sampler z axis.The fse then validated the analyzer by running (qc) with results within acceptable range.The aia-2000 analyzer is operating as expected.No further action required by field service.A 13-month complaint history review and service history review through aware date of event for similar complaints was performed for "imprecision quality control results for beta human chorionic gonadotropin (bhcg) and human chorionic gonadotropin (hcg)" on the aia-2000 analyzer.There were no similar complaints found during the searched period.The analyte application manual for beta human chorionic gonadotropin (bhcg) states the following: evaluation of results quality control in order to monitor and evaluate the precision of the analytical performance, it is recommended that commercially available control samples be assayed daily.The minimum recommendations for the frequency of running internal control material are: ¿ after calibration, two levels of controls are run in order to accept the calibration curve.¿ the two levels of controls are also repeated after calibration when certain service procedures are performed (e.G.Temperature adjustment, sampling mechanism changes, maintenance of the wash probe or detector lamp adjustment or change).¿ after daily maintenance, at least two levels of the control should be run in order to verify the overall performance of the tosoh aia system analyzers.If one or more control sample value(s) is out of the acceptable range, it will be necessary to investigate the validity of the calibration curve before reporting patient results.Standard laboratory procedures should be followed in accordance with the regulatory agency under which the laboratory operates.The analyte application manual for human chorionic gonadotropin (hcg) states the following: evaluation of results quality control in order to monitor and evaluate the precision of the analytical performance, it is recommended that commercially available control samples be assayed daily.The minimum recommendations for the frequency of running internal control material are: after calibration, two levels of the internal control are run in order to accept the calibration curve.The two levels of controls are repeated when certain service procedures are performed (e.G.Temperature adjustment, sampling mechanism changes, maintenance of the wash probe, or detector lamp adjustment or change).After daily maintenance, two levels of the control shall be run in order to verify the overall performance of the tosoh aia system analyzer.If one or more control sample value(s) is out of the acceptable range, it will be necessary to investigate the validity of the calibration curve before reporting patient results.Standard laboratory procedures should be followed in accordance with the regulatory agency under which the laboratory operates.The most probable cause was due to a lubrication problem with the main arm sampler z axis, component failure.
|