| Model Number |
22800810102011010 |
| Medical Device Problem Code |
Failure to Power Up (1476)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Correction: d2.The device was returned to zoll medical corporation for evaluation.During investigation of the device to determine root cause, the technician observed physical damages to the on/off power button gasket and switch consistent with mis-handling.The on/off power button was observed torn during evaluation.The on/off gasket and main printed circuit board (pcb) were replaced.The device was recertified and returned to the customer.Analysis of reports of this type has not identified an increase in trend.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during incoming inspection, the device would not power up.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has not received the product for evaluation and this complaint is still under investigation.
|
| |
|