| Model Number |
30001004101110013 |
| Medical Device Problem Code |
Output below Specifications (3004)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has not received the device for evaluation and this complaint is still under investigation.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during functional testing, the device's defib output was out of specification.Complainant did not indicate that there was any patient involvement in the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction: d2.The customer was contacted for return of the suspect product.The product has not been returned to zoll for evaluation.
|
| |
|
Search Alerts/Recalls
|