| Model Number |
DDMB1D1 |
| Medical Device Problem Code |
Electromagnetic Interference (1194)
|
| Health Effect - Clinical Codes |
Dizziness (2194); Presyncope (4410); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/15/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient presented with dizziness and near syncope.The right ventricular (rv) lead and right atrial (ra) lead exhibited oversensing with electromagnetic interference observed on the implantable cardioverter defibrillator (icd).Additionally, the ra lead exhibited undersensing during atrial tachycardia/atrial fibrillation (at/af) stored episodes resulting in termination of detected events in progress.The icd, ra lead, and rv lead remain in use.No patient complications have been reported as a result of this event.
|
| |
|
Search Alerts/Recalls
|