| Catalog Number |
519650 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/21/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
According to the available information, the anchor on the static side pulled away from the mesh after successful implantation when normal pressure was applied to load the other side.A new altis sling was successfully implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
The lot number was reviewed for complaint trend, non-conforming report and capa.Devices met specification prior to release.No trends were noted for complaints and there were no non-conforming reports or capas that were confirmed in veeva to be associated.The device was not returned for evaluation.Without the benefit of analyzing the device, coloplast cannot comment on the condition of the device.Should additional information prompt us to alter or supplement any information or conclusions contained in this report, a follow-up report will be submitted.
|
| |
|