| Model Number |
Mi1200 SYNCHRONY PIN |
| Medical Device Problem Code |
Impedance Problem (2950)
|
| Health Effect - Clinical Code |
Head Injury (1879)
|
| Date of Event |
05/04/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The user's hearing performance is affected after multiple traumas to the head.Reimplantation is considered.
|
| |
|
Additional Manufacturer Narrative
|
|
The device has not been explanted.If it should be explanted, it should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
|
| |
|
Event or Problem Description
|
|
The user's hearing performance is affected after multiple traumas to the head.The user has been re-implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Conclusion: damage to the reference electrode, likely caused by the repeated external impacts, was determined to be the root cause of device failure.Other mechanical damages found are attributable to the removal surgery.This is a final report.
|
| |
|