Model Number 436-0478
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Codes Failure of Implant (1924); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/10/2026
Type of Reportable Event Malfunction
Event or Problem Description
A patient underwent a 4-level anterior cervical discectomy and fusion procedure on (b)(6) 2025.About 3 months postoperatively, radiographic images revealed a screw backout at the most inferior level.There are no plans for revision surgery.
 
Additional Manufacturer Narrative
The device has not returned for evaluation.Radiographic images were provided which confirm the event of an inferior screw backout.A review of the device history records could not be performed as the identifying lot number was not provided.Based on the information provided, the root cause could not be determined.Alphatec spine is submitting this report to comply with the fda regulations 21 cfr 803.Alphatec spine has made reasonable efforts to provide as much relevant information as available.Any field that is left blank was not known at the time of this submission.If additional information is provided, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INSIGNIA ANTERIOR CERVICAL PLATE SYSTEM
Common Device Name
APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY
Manufacturer (Section D)
ALPHATEC SPINE, INC.
1950 camino vida roble
carlsbad CA 92008
Manufacturer (Section G)
ALPHATEC SPINE, INC.
1950 camino vida roble
carlsbad CA 92008
Manufacturer Contact
michael hendrickson
1950 camino vida roble
carlsbad, CA 92008
7603566777
MDR Report Key25137441
Report Number2027467-2026-00089
Device Sequence Number9543260
Product Code KWQ
UDI-Device Identifier00190376491370
UDI-Public(01)00190376491370
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
K213443
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 05/11/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number436-0478
Device Catalogue Number436-0478
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/14/2026
Initial Report FDA Received Date05/11/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-