• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS, INC. BIPAP A40; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. BIPAP A40; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE Back to Search Results
Model Number GBX3100S19
Medical Device Problem Code Electrical /Electronic Property Problem (1198)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/30/2026
Type of Reportable Event Malfunction
Event or Problem Description
The manufacturer received information alleging a ventilation inoperative error code was found on a bipap a40 pro device.The device was not in use on a patient at the time of the occurrence.During the evaluation it was noted that an error code, defective defective electrically erasable programmable read-only memory., was observed on the device's error log.There was no documentation by the third-party service technician that stated on a root cause and/or any component replacement to resolve the failure of the outlet pressure sensor error code.A final report is being submitted.If new information becomes available following the completion of the device repair that changes the outcome of the investigation and associated medical device reporting a supplemental report will be completed.The root cause isn't confirmed at this time.Additionally, during the service of the device, two additional, error codes, high internal o2 high alarm and fan status, were observed on the device's error log.There was no documentation by the third-party service technician that stated on a root cause and/or any component replacement to resolve the high internal o2 high alarm and fan status error codes.These two error codes were unrelated to the reportable issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIPAP A40
Common Device Name
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer Contact
melissa rosko
1001 murry ridge lane
murrysville, PA 
7247330200
MDR Report Key25137706
Report Number2518422-2026-012131
Device Sequence Number9779709
Product Code MNT
UDI-Device Identifier00606959055063
UDI-Public(01)00606959055063
Combination Product (Y/N)N
Initial Reporter CountryEI
PMA/510(K) Number
K121623
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Y
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Recall
Type of Report Initial
Report Date (Section B) 05/11/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberGBX3100S19
Device Catalogue NumberGBX3100S19
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/30/2026
Initial Report FDA Received Date05/11/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured11/21/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-1813-2024
Patient Sequence Number1
Patient SexUnknown
-
-