| Model Number |
1054655B |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/12/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A ventilator was returned to a manufacturer's service center for preventive maintenance.There was no harm or injury reported.The device was not in patient use.During the evaluation a failure of the active exhalation control module to close was observed.The device's active exhalation control module was replaced to address the issue.The device was cleaned and tested and passed all final testing.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported failure of active exhalation control module to close is accommodated in the product risk management file as being linked to hazard with description patient exposure to function / deterioration of function.Complaints for this failure are reviewed via the post market complaint trending and escalation processes to assess for the observed probability levels and compare them with the predicted levels in the risk file for the hazards linked to the failure.As of the date of submission for this report, there is no indication that complaints for the failure in question has led to unacceptable increase in probability levels for the hazardous situations in scope, and therefore no further action will be pursued.Qa continues to monitor complaints for this issue per the cadence in the complaint trending procedures.Dhr review was performed for the complaint device sn.Review confirms that the device met the final acceptance criteria and was released for final distribution.
|
| |
|