| Model Number |
GIF-2TH180 |
| Medical Device Problem Code |
Residue After Decontamination (2325)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/03/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for inspection.Based on the results of the investigation, the definitive root cause of the reported issue could not be determined.However, it is likely the use of products that contain silicone led to the malfunction.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|
Event or Problem Description
|
|
It was observed during the device evaluation that the gastrointestinal videoscope had white foreign material in the air/water cylinder and air/water channel.There was no patient involvement.
|
| |
|