| Model Number |
SR-200-NPS |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Stroke/CVA (1770)
|
| Date of Event |
04/15/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the patient experienced a stroke after the procedure.The patient presented for a transcarotid revascularization (tcar) procedure.An enroute transcarotid neuroprotection system was selected for use.The patient experienced a stroke of unknown etiology after the procedure.No further information was provided despite request by boston scientific (bsc).
|
| |
|
Additional Manufacturer Narrative
|
|
B3: date of event: the event date was approximated to the date bsc became aware of the event, 04/15/2026.Detailed product information was not provided to bsc.Based on the nature of the information provided to bsc, it is not possible to perform a good faith effort to obtain additional information.Because the product is unknown at this time, we are unable to provide the complete unique identifier (udi) # and other product specific information.If additional details become available, a supplemental report will be submitted.Of note, bsc was made aware of 8 events from the same anonymous physician survey.The corresponding bsc complaint id's are as follows: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that the patient experienced a stroke after the procedure.The patient presented for a transcarotid revascularization (tcar) procedure.An enroute transcarotid neuroprotection system was selected for use.The patient experienced a stroke of unknown etiology after the procedure.No further information was provided despite request by boston scientific (bsc).
|
| |
|
Additional Manufacturer Narrative
|
|
B3 - date of event: the event date was approximated to the date bsc became aware of the event, 04/15/2026.Detailed product information was not provided to bsc.Based on the nature of the information provided to bsc, it is not possible to perform a good faith effort to obtain additional information.Because the product is unknown at this time, we are unable to provide the complete unique identifier (udi) # and other product specific information.If additional details become available, a supplemental report will be submitted.Of note, bsc was made aware of 8 events from the same anonymous physician survey.The corresponding bsc complaint id's are as follows: (b)(4).Updated fields: h6 - evaluation method codes; evaluation conclusion codes.Review of the instructions for use (ifu) confirmed that the content was sufficient and did not contribute to the reported event; therefore, no updates are required to the document at this time.A review of the enroute transcarotid neuroprotection system device confirmed that the reported event is a known event defined in the product's risk management documentation.This event type has been accounted for during product risk analysis to support acceptable risk benefit for the product.Based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of known inherent risk of device.
|
| |
|