| Catalog Number |
72202901 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/20/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H11: internal complaint reference: (b)(4).H3, h6: this complaint was opened by smith+nephew to document a product problem associated with a smith+nephew device.The reported problem relates to known risks with the device itself or with its use that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are submitting this report in accordance with applicable regulations.If additional investigative findings or information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
|
| |
|
Event or Problem Description
|
|
It was reported that during surgery, one (1) footprint ultra anchor was fractured when it was being implanted.The procedure was performed using an equivalent s+n back-up.It is unknown if there was any surgical delay.No further complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: internal complaint reference: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that during a rotator cuff repair, one (1) footprint ultra anchor was fractured when it was being implanted.The fractured pieces were removed using tweezers.The procedure was performed, after a non-significant delay, using an equivalent s+n back-up in the original drilled bone hole.No further complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
H11 h3, h6 the reported device was received for evaluation.A visual inspection showed the device was returned in original packaging with the batch number of the complaint on the label.The anchor assembly was returned with the distal tip of the anchor body fractured.The anchor plug has no visible defect.The insertion device and green retention suture have no visible defect.No other suture was returned.An additional assessment found that based on the condition of the product material found during visual inspection, it was determined that additional material testing is not required.A device deficiency was not identified, and the root cause of the reported event could not be determined since the device was not returned for evaluation.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A review of the material characteristics found that the raw material strength requirements and storage requirements are specified and each lot must be accompanied by a material certificate of analysis.A definitive root cause for the reported issue could not be determined.Factors that could have contributed to the failure include procedural variance, or deviations from established protocols, increasing risks and leading to unintended consequences.Other factor lines include application of improper/excessive force.Based on this investigation, the need for corrective action is not indicated.
|
| |
|