Block h6: imdrf device code a150201 captures the reportable event of stent failure to deploy.Block h11: following an update to the axios risk documentation, this event is being reported as part of the efforts associated with boston scientific's nonconforming events and prevention investigation, (b)(4).Investigation results: based on the available information, boston scientific could not confirm the reported event of stent failure to deploy and handle break.The device was not returned for analysis; therefore, a technical analysis could not be performed.There is not enough evidence to determine whether the reported events were due to the physician's technique during the procedure or was related to a device malfunction.Good faith effort attempts were made to try and retrieve the device; however, response was not received.Device technical analysis: the device was not returned for analysis; therefore, a technical analysis could not be performed.Device history record review: it was confirmed this device met manufacturing specifications prior to distribution and there were no manufacturing deviations which could have contributed to the reported event.Risk review: a risk review was completed and confirmed that the event of "stent failure to deploy" was defined in the risk documentation and is documented accordingly in the prr.This event type has been accounted for during product risk analysis to support acceptable risk benefit for the product.Investigation conclusion: based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of cause not established.
|