U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION AXIOS?; UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION AXIOS?; UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES) Back to Search Results
Model Number M00553640
Medical Device Problem Codes Break (1069); Positioning Failure (1158)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/10/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported to boston scientific corporation that an axios stent and electrocautery enhanced delivery system was intended to be implanted in the stomach to gather a fluid collection during an endoscopic ultrasound (eus) performed on (b)(6) 2024.During the procedure, the physician reported that the axios stent was unable to deploy.In order to facilitate the release, the physician tried to redeploy it but did not work.The handle broke so the physician had to remove the device.The stent was removed fully covered by the outer sheath and the procedure was able to be completed by using a different device.There were no patient complications reported as a result of this event.
 
Additional Manufacturer Narrative
Block h6: imdrf device code a150201 captures the reportable event of stent failure to deploy.Block h11: following an update to the axios risk documentation, this event is being reported as part of the efforts associated with boston scientific's nonconforming events and prevention investigation, (b)(4).Investigation results: based on the available information, boston scientific could not confirm the reported events of stent failure to deploy and handle break.The device was not returned for analysis; therefore, a technical analysis could not be performed.There is not enough evidence to determine whether the stent failure to deploy and handle break was due to the physician's device manipulation during the procedure or related to a device malfunction.The sample is unavailable, disposed on customer site.Device technical analysis: the complaint device was not returned; therefore, product analysis could not be performed.Device history record review: it was confirmed this device met manufacturing specifications prior to distribution and there were no manufacturing deviations which could have contributed to the reported event.Risk review: a risk review was completed and confirmed that the event of stent failure to deploy was defined in the risk documentation.This event type has been accounted for during product risk analysis to support acceptable risk benefit for the product.Investigation conclusion: based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of cause not established.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AXIOS?
Common Device Name
UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
christina lam
300 boston scientific way
marlborough, MA 01752
9145716918
MDR Report Key25144866
Report Number3005099803-2026-01824
Device Sequence Number21176908
Product Code KNS
UDI-Device Identifier08714729904588
UDI-Public(01)08714729904588(17)250621(10)31870383
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
K233318
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 05/11/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/21/2025
Device Model NumberM00553640
Device Catalogue NumberM00553640
Device Lot Number0031870383
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/15/2024
Initial Report FDA Received Date05/11/2026
Date Device Manufactured06/22/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexMale
-
-