Model Number M00553660
Medical Device Problem Code Positioning Failure (1158)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/01/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Block d4, h4: the complainant was unable to provide the complaint device lot number.We are unable to provide the device manufacture date.If additional details become available, a supplemental report will be submitted.Block h6: imdrf device code a150201 captures the reportable event of stent failure to deploy.Block h11: following an update to the axios risk documentation, this event is being reported as part of the efforts associated with boston scientific's nonconforming events and prevention investigation, (b)(4).Investigation results: based on the available information, boston scientific could not confirm the reported event of stent failure to deploy.The device was not returned for analysis; therefore, a technical analysis could not be performed.There is not enough evidence to determine whether the stent failure to deploy was due to the physician's device manipulation during the procedure or related to a device malfunction.The sample is unavailable, disposed on customer site.Device technical analysis: the complaint device was not returned therefore, product analysis could not be performed, for this reason and due to the lack of evidence is unable to confirm the reported event.Device history record review: it was confirmed this device met manufacturing specifications prior to distribution and there were no manufacturing deviations which could have contributed to the reported event.Risk review: a risk review was completed and confirmed that the event of "stent failure to deploy" was defined in the risk documentation.This event type has been accounted for during product risk analysis to support acceptable risk benefit for the product.Investigation conclusion: based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of cause not established.
 
Event or Problem Description
It was reported to boston scientific corporation that an axios stent and electrocautery enhanced delivery system was intended to be implanted transgastric to treat a pseudocyst during a procedure performed on (b)(6) 2024.During the procedure, the physician reported that the flange would not open when attempting to deploy.Additionally, the physician tried multiple times to deploy stent but was unsuccessful.The stent was removed fully covered by the outer sheath and the procedure was able to be completed by replacing and using another of the same device.There were no patient complications reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AXIOS STENT AND ELECTROCAUTERY ENHANCED DELIVERY SYSTEM
Common Device Name
PANCREATIC STENT, COVERED, METALLIC, REMOVABLE
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
christina lam
300 boston scientific way
marlborough, MA 01752
9145716918
MDR Report Key25144869
Report Number3005099803-2026-01820
Device Sequence Number18486501
Product Code PCU
UDI-Device Identifier08714729951179
UDI-Public(01)08714729951179
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
K220112
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 05/11/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberM00553660
Device Catalogue NumberM00553660
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/01/2024
Initial Report FDA Received Date05/11/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-