• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HOYA CORPORATION PENTAX TOKYO OFFICE PENTAX MEDICAL; VIDEO GASTROSCOPE (US ONLY)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HOYA CORPORATION PENTAX TOKYO OFFICE PENTAX MEDICAL; VIDEO GASTROSCOPE (US ONLY) Back to Search Results
Model Number EG29-I10C
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/05/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The product was returned to pentax medical for repair.Our technician checked the returned unit and confirmed that the ccd drive pcb fluid damage.Based on the result, we concluded that it was caused due to the fluid damage from the ccd drive pcb.In addition, our technician confirmed that the ccd drive pcb corroded; however, this defect is not the main cause, and/or irrelevant to the alleged complaint.Based on the technical report "hr-rpt-0586(image failure)" and/or the risk analysis results, it was evaluated to submit mdr.
 
Event or Problem Description
The time of event is unknown.There was no report of patient harm.Video image failure(fluid damage).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENTAX MEDICAL
Common Device Name
VIDEO GASTROSCOPE (US ONLY)
Manufacturer (Section D)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishima-shi, tokyo 196-0 012
JA  196-0012
Manufacturer (Section G)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishima-shi, tokyo 196-0 012
JA   196-0012
Manufacturer Contact
paresh patel
3 paragon drive
montvale, NJ 07645
8004315880
MDR Report Key25146003
Report Number9610877-2026-50991
Device Sequence Number16140290
Product Code FDS
UDI-Device Identifier04961333237586
UDI-Public(01)04961333237586(11)190701
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
K190805
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/12/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberEG29-I10C
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/05/2026
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/05/2026
Initial Report FDA Received Date05/12/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/01/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-