| Model Number |
01.18.132 |
| Medical Device Problem Code |
Osseointegration Problem (3003)
|
| Health Effect - Clinical Codes |
Inadequate Osseointegration (2646); Aseptic loosening (4883)
|
| Date of Event |
04/21/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Batch review performed on 22 april 2026 lot 165809: (b)(4) items manufactured and released on 16-jan-2017.Expiration date: 2022-01-04.No anomalies found related to the problem.To date,
(b)(4) items of the same lot have been sold with no similar reported events during the period of review.Root cause: aseptic loosening is possible literature described adverse event after primary hip arthroplasties and causes are often unknown.There is no indication that any potential issue with the device may have caused or contributed to the event and the investigation does not indicate any potential manufacturing related issue or systemic deficiency.
|
| |
|
Event or Problem Description
|
|
At about 9 years after the primary, the patient came in presenting pain due to a loose stem and the cause is unknown.There were no indications of trauma or fall.The surgeon revised the stem and head.The surgery was completed successfully.
|
| |
|
Search Alerts/Recalls
|