• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SYNTHES PRODUCTS LLC; DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SYNTHES PRODUCTS LLC; DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) Back to Search Results
Catalog Number EMAX2PLUS
Medical Device Problem Code Fail-Safe Did Not Operate (4046)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11 additional narrative: a visual and functional assessment was performed on the device.The following steps failed: 2.3.1 visual assessment 2.3.2 visual assessment 2.5.3 cable assessment 2.13.1 safety assessment.During the service evaluation the following failures were identified: visual : cord damaged visual : breaded stainless steel wire tether damage/came off functional : moving parts do not move smoothly functional : device runs in locked position.Conclusion: failure reported is confirmed root cause: component failure from wear.
 
Event or Problem Description
It was reported that during service and evaluation, it was determined that the motor device ran in the locked position.It was noted in the service order that the device had an unspecified malfunction.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of the event was unknown.However, it was reported that the event occurred in 2026.All available information has been disclosed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NA
Common Device Name
DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)
Manufacturer (Section D)
DEPUY SYNTHES PRODUCTS LLC
4500 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
elba bello
4500 riverside drive
palm beach gardens, FL 33410
9083863534
MDR Report Key25150301
Report Number1045834-2026-00108
Device Sequence Number9213243
Product Code HBE
UDI-Device Identifier00845384002055
UDI-Public(01)00845384002055
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K080802
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberEMAX2PLUS
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/22/2026
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/07/2026
Initial Report FDA Received Date05/12/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/11/2014
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-