• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GIVEN IMAGING LTD., YOQNEAM BRAVO; ELECTRODE, PH, STOMACH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GIVEN IMAGING LTD., YOQNEAM BRAVO; ELECTRODE, PH, STOMACH Back to Search Results
Model Number FGS-0636
Medical Device Problem Code Positioning Failure (1158)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 04/14/2026
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
D10 concomitant product: fgs-0636, fgs-0636 cf delivery dev caps bravo x1 (lot# 65922f).H3: evaluation summary: medtronic conducted an investigation based upon all information received.The capsule and delivery system were returned for evaluation.Visual inspection of the delivery device and capsule found no notable conditions.It was reported that the bravo capsule failed to attach to patient's esophagus.The reported issue was confirmed.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
It was reported that the first capsule was deployed with a suction time of 1 minute and the vacuum pump pressure reached 550 mmhg using suction and the capsule failed to attach to the esophagus and dropped into the patient¿s stomach.The capsule was retrieved from the stomach using a roth net.The same issue occurred with a second capsule and also dropped into the patient¿s stomach.A third capsule was then used, and the suction time was extended to 2 minutes, after which the third capsule was successfully deployed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BRAVO
Common Device Name
ELECTRODE, PH, STOMACH
Manufacturer (Section D)
GIVEN IMAGING LTD., YOQNEAM
yetsira 13 street
yoqneam 20692
IS  20692
Manufacturer (Section G)
GIVEN IMAGING LTD., YOQNEAM
yetsira 13 street
yoqneam 20692
IS   20692
Manufacturer Contact
mike bocchino
8200 coral sea st ne
mounds view, MN 55112-4391
8006464633
MDR Report Key25150880
Report Number9710107-2026-00208
Device Sequence Number10733549
Product Code FFT
UDI-Device Identifier07290101369714
UDI-Public(01)07290101369714(17)261216(10)65922F(20)01
Combination Product (Y/N)N
Initial Reporter CountrySN
PMA/510(K) Number
K102543
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 05/12/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberFGS-0636
Device Catalogue NumberFGS-0636
Device Lot Number65922F
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/27/2026
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/15/2026
Initial Report FDA Received Date05/12/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/16/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
PLEASE SEE NOTE ON H11.
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
-
-