MEDTRONIC MEXICO S. DE R.L. DE CV EVOLUT FX PLUS VALVE; AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIV
|
Back to Search Results |
|
| Model Number |
EVFXPLUS-34 |
| Medical Device Problem Code |
Material Invagination (1336)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/11/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that during the attempted implant of a transcatheter aortic valve, upon the first deployment attempt at a target implant depth of 2 millimeters (mm) on the native annulus, an infold was identified.Subsequently, the valve was recaptured, and the system was withdrawn from the patient.It was noted that \"echmo\" was provided as a treatment.Subsequently, a new valve and delivery catheter system (dcs) were opened and prepared.Following insertion into the patient, it was observed that the correct positioning of the valve could not be achieved.4 deployment attempts were made, however correct positioning was not achieved on any of the 4 attempts.Subsequently, the valve was recaptured for a final time, and the system was withdrawn from the patient.Per the physician, the patient's native bicuspid valve contributed to the infold.
|
| |
|
Additional Manufacturer Narrative
|
|
Continuation of d10: product id {d-evolutfx-2329}; product lot/serial number {(b)(6)}; product type: {0195-heartvalves}; implant date {non-implantable} medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Additional Manufacturer Narrative
|
|
Updated data: b5.A second paragraph was added.E.Initial reporter name corrected data - inadvertently omitted from the initial medwatch 3500a: b2.Outcome attributed to adverse event d10.Concomitant device: product id: d-evolutfx-34, lot #: 0013259693 erroneously incorrect on the initial medwatch 3500a.Additional codes.Annex f code f1203 replaced f12.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that during the attempted implant of a transcatheter aortic valve, upon the first deployment attempt at a target implant depth of 2 millimeters (mm) on the native annulus, an infold was identified.Subsequently, the valve was recaptured, and the system was withdrawn from the patient.It was noted that \"echmo\" was provided as a treatment.Subsequently, a new valve and delivery catheter system (dcs) were opened and prepared.Following insertion into the patient, it was observed that the correct positioning of the valve could not be achieved.Four deployment attempts were made; however, correct positioning was not achieved on any of the 4 attempts.Subsequently, the valve was recaptured for a final time and the system was withdrawn from the patient.Per the physician, the patient's native bicuspid valve contributed to the infold.Additional information was received to report the reference to \"echmo\" treatment was a spelling error and meant to indicate extracorporeal membrane oxygenation (ecmo).
|
| |
|