On-site investigation was performed by our field service engineer.The claimed issue was reproduced during troubleshooting.According to received information in service report, replacement of the drainage valve solved the issue.The ventilator passed all functional and safety tests and was cleared for clinical use.The issue has been resolved and the device has been returned to use in good working condition.The drainage valve is powered with 12 v and electrically controlled by the pc board.The purpose of the drainage valve is to remove water collected in the water separator at regular intervals.The valve is closed in non-activated condition.During operation, the valve will be activated (open) for approx.0.5 seconds twice every minute by the timing circuit on the pc board.When the valve opens, the water collected in the water separator will be transported to the drainage bottle.The faulty drainage valve may lead to moisture air entering the air gas module in the connected ventilator and may lead to stop of ventilation.Water in the air gas module may damage the air gas module.The part has not been available for the further analyze of the issue.The cause of the issue was related to drainage valve.This event occurred on the hong kong market on a significantly similar device to ¿compressor mini 115v 60hz¿ which is sold in the us.Therefore, for d4 primary di number has been used from compressor mini 115v 60hz.Provided in report: d4 serial # and manufacture date correspond to device involved in the event, which is "compressor mini 230v¿.
|