TORNIER INC TORNIER PERFORM ANATOMIC AUG SMALL RIGHT KEELED 15 DG; PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED
|
Back to Search Results |
|
| Catalog Number |
DWK503RA15 |
| Medical Device Problem Codes |
Off-Label Use (1494); Osseointegration Problem (3003)
|
| Health Effect - Clinical Codes |
Discomfort (2330); Inadequate Osseointegration (2646); Muscle/Tendon Damage (4532); Radiolucent Lines (4848)
|
| Date of Event |
01/20/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The reported event could be confirmed since evidence was provided, based on images and reported information.The returned device inspection revealed the following: the identification of the device is confirmed.Bone and soft tissue residues are observed on the device.Surfaces of the device are degraded scratches, pitting marks, deformed material and edges.Since images and information were provided, the opinion of a medical expert was sought and stated as follows both the humeral and glenoid components are well-positioned.This was also mentioned in the notes of the complaint that humeral implants were well fixed.No specific factors, especially no decentering of the humeral component relative to the glenoid.Consequently, no clear radiological signs of rotator cuff dysfunction.There is a radiolucent line around at the glenoid component-bone interface.Radiologically no absolute signs of loosening, no migration, intraoperatively it was mentioned, however.No unintended humeral head-glenoid mismatch on the x-ray.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.No indications of material, manufacturing or design related problems were found during the investigation.A review of the labeling did not indicate any abnormalities.According to the humeral head-glenoid combinations recommended in the current tornier perform anatomic augmented glenoid operative technique, the head-glenoid combination in this case is within an unacceptable mismatch range, combination of a 50 into16 size humeral head with a small size glenoid.Based on the investigation, the root cause was attributed to a user related issue.The failure was caused by an off-label use, unacceptable head-glenoid combination.If any additional information is provided, the investigation will be reassessed.
|
| |
|
Event or Problem Description
|
|
It was reported that, the patient underwent a right primary anatomic shoulder replacement on (b)(6) 2024.Advised by surgeon yesterday that the patient now requires a revision (removal of all implants), and conversion to a reverse shoulder replacement, as the patient's rotator cuff has failed.The surgery was scheduled for (b)(6) 2026.Right primary shoulder arthroplasty on (b)(6) 2024.Patient had been complaining of discomfort in prosthetic right shoulder joint approx 6 months post-op.X-ray at this time showed some evidence of lucency around glenoid component.Ct scan at 12 months post-op showed increasing lucency.He complained of more discomfort.Ct scan late 2025 showed loose glenoid component, revision surgery booked.Simpliciti head removed, simpliciti nucleus was solid, but the revision tools got it out quite clean.Minimal bone loss.Glenoid perform plus poly was very loose, cavity behind it.Component had minimal cement still adhered to it.Multiple tissue samples taken, though surgeon stated that there were no clinical signs of infection.Copious lavage.Shoulder joint reduced, stable through full range of motion.Explanted prosthesis being returned (in a steri-peel pkt) with loan instruments.
|
| |
|
Search Alerts/Recalls
|
|
|