BIOSENSE WEBSTER INC VIZIGO; INTRODUCER, CATHETER Back to Search Results
Catalog Number D138504
Medical Device Problem Code Backflow (1064)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/20/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a patient underwent a cardiac ablation procedure that involved usage of vizigo sheaths.However, the first two sheaths had the same issue: air drawn into the side port.During the first insertion of the varipulse catheter into the sheath, air was continuously drawn into the side port.When the sheath was replaced, the same issue occurred with the second sheath.The issues did not resolve despite the use of an air tray.Both sheaths had been replaced and the procedure was completed using a third sheath.No air was drawn into the patient's body.No patient consequences were reported.There are two reports for the first two vizigo sheaths.This report is for lot 60000681.
 
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by biosense webster, inc., or its employees that the report constitutes an admission that the product, biosense webster, inc.Or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Manufacturer's reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VIZIGO
Common Device Name
INTRODUCER, CATHETER
Manufacturer (Section D)
BIOSENSE WEBSTER INC
31 technology drive, suite 200
irvine CA 92618
Manufacturer (Section G)
FREUDENBERG MEDICAL LLC
2301 centennial blvd
jeffersonville IN 47130
Manufacturer Contact
henry boden
31 technology dr
irvine, CA 92618
9497882078
MDR Report Key25152909
Report Number2029046-2026-01549
Device Sequence Number21135622
Product Code DYB
UDI-Device Identifier10846835026252
UDI-Public(01)10846835026252(17)270422(10)60000681
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K170997
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/16/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberD138504
Device Lot Number60000681
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/20/2026
Supplement Date Received by Manufacturer05/22/2026
Initial Report FDA Received Date05/12/2026
Supplement Report FDA Received Date06/16/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/29/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1